:: Jejoong-Medical ::
 
HOME Product Zatha
ZATHA-S
ZATHA-S
ZATHA-S
ZATHA-S
ZATHA-S
ZATHA-S
File ZATHA_manual(English).pdf (442.5K) Last Update : 2012.04.09
File (No data.)

Main Function
Zatha series use a principle that works cutting and coagulation of cellular tissue via high-frequency current through the body contact load and resistance by using the heat generated.
As the high-frequency current through the body, heat will be occurred by heat and arc discharge in active electrode and return-pad fitted wide area in the body. So, the heat makes heating the cellular tissue and it is possible to cut and coagulate as operation.
Return-pad and the active electrode are mono-polar in combination. In another function, the end of electrode as shape of forceps is a pair of electrode. The high frequency current works among the electrode in the forceps. It called a bipolar.

Feature
Safety
In the case, the connection is incorrect between patients and attached return-pad with a patient as operation, automatically RF output will be blocked and sound an alarm.

Preciseness
All the operations of ZATHA Series are digitalized output 1W of unit by means of a microprocessor. So, more precise operation is possible. Also user turns off a unit, the unit memories final setting value in last operation.

Specification
ZATHAZATHA-S
Electrical specification
Supply voltage100~230V100~230V
Frequency50/60Hz50/60Hz
Power input during max.900VA1100VA
FuseT10AT10A
Power output specification
Maximum CUT200W300W
Maximum BLEND150W200W
Maximum COAG150W200W
Maximum BIPOLAR70W100W
Dimension and weight
Width x Depth x Height330x350x130mm330x350x130mm
Net Weight7.5kg7.5kg